Caution
Read the following items carefully before
operating RK2671 series voltage withstand test instrument.
1. Use three-hole power supply plug and insure GND connection well.
2. GND port and power supply GND port should be
connected well.
3. Wear insulation glove, stand on insulation mat.
4. Anti-clockwisely circumvolve “V-ADJ”
rotary switch to the
maximum before turn on test instrument.
5. Before preset all functions, reset instrument.
6.
Don’t
connect test port and test wire when doing test.
7. Don’t
connect HV output port with shield box or GND wire to
avoid instrument damage.
8. Turn off power when accident take place.
9. Examine and repair when Indicator or annunciator is
out of function.
10. Operate carefully to avoid hazard when take the remote test.
11.The instrument must be placed on the worktable of conductive
materials.
12.The operator can not wear clothes with metal decoration or wear
metal accessories.
13.When the tester is in the test conditions, the test line, the test
probes, output end , and the test
objects are
all with high pressure, do not touch!
|