1. Before using the tank, it is necessary to
add liquid medium, otherwise, it will damage the heater. The medium
level must be lower than the platform.
2. The selection of the liquid medium in the
tank shall be in accordance with the following principles:
A. When the operating temperature is 5℃-85℃,
the liquid medium is usually water
B. When the operating temperature is 85℃-95℃,
the liquid medium can be 15% glycerin-water solution
C. When the operating temperature is
lower than 5℃,
the liquid medium is generally alcohol or freezing liquid
3. In
order to prolong the service life of the refrigeration system, the
system adopts startup protection. Once the refrigeration system
stops working, it will need 10-20minutes' space to turn on the
switch again (for example, when the switch is turned on to start the
compressor in less than 10 minutes from the last operation, it is a
normal phenomenon that the system will not start.)
4. Power supply should be greater than the
total power of the machine, the power supply must have a good
grounding device.
5. The machine should be placed in a dry and
ventilated place.
6. When the working temperature is very low,
please do not put your hands into the tank to avoid frostbite.
7. Cut off power supply when operation is
finished.
. |